- Storage Space: Two spacious drawers to hold essential anesthesia kit and accessories
- Cylinder Yokes: Color coded gas specific (pin indexed) yokes with sliding stainless steel clamping bars for easy handling.
- Holding 2 (Two) for O₂ and 1 (One) for N₂O, 5 Ltr. water capacity cylinders
- Non-return / Pressure Relief Valve: Minimises risk of back flow of gases, blows off whe/n pressure exceeds 125 Cm H₂O
- Regulators: Designed for higher efficiency, long service life and reliability for O₂ and N₂O
- Oxygen Flush: Emergency oxygen flush switch at table level to give 30–70 Ltrs/min for emergencies
- O₂ Failure Warning Alarm: Self activated electro-pneumatic audio visual warning device to indicate drop in O₂ pressure below 2.5 Bar
- Hypoxic Guard (Mechanical): Mixing device to proportion the N₂O in such a way that minimum of 25% O₂ is always delivered in the gas mixture
- Common Gas Outlet (CGO): Conveniently located CGO with ISO tapers for 22 & 15 mm connections
- Basal flow with On/off Switch: 250 ± 50 ml basal flow of Oxygen as soon as machine is switched on (optional)
- Pipe Line Gas Inlets: Non-Interchangeable Screw Threaded (NIST) pipeline inlets one each for O₂, N₂O and Air.
- In case of drop in piped gas pressure the pin index cylinders take over automatically
- Pressure Gauges: Large diameter, color coded gauges fitted at convenient angle for better visibility and clarity.
- One for each of the pin indexed yokes and one each corresponding to piped gas inlet
- Selectatec Brackets: Twin Selectatec Block
- Breathing Circuits: Bains Circuit & Circle Absorber Y-Circuit provided as standard
- Communication port: RS232 & LAN Port
- Cascaded 6 tube flow meter
- Twin Selectatec mounting for agent specific temperature and pressure compensated vaporizers
- NIST inlets for piped medical gases for O₂ , N₂O & Air as standard
- Single Bi-stable system ON / OFF switch for gases & power supply
- Pull out writing table with folding lock and Common gas outlet (CGO)
- Open/Close change over switch
- Suitable for open, semi-open & closed circuit system application
- Electrical outlets for running other medical equipment
- Auxiliary O₂ outlet with flowmeter as standard
- Rotameter backlit illumination
- Supports Low-Flow Anesthesia